Document Type : Research Paper
Authors
1 Materials and Energy Research Center, Tehran, Iran,
2 Department of Clinical Sciences, Surgery and Radiology Section, Faculty of Veterinary Medicine, University of Tehran, Tehran-Iran,
3 Department of Biomedical Engineering, Science and Research Campus, Islamic Azad University, Tehran, Iran,
4 Laboratoire de Métallurgie Physique et Génie des Matériaux, Université des Sciences et Technologies de Lille, UMR CNRS 8517, Lille, France
Abstract
Keywords
1. Introduction
Titanium has been widely used in orthopedic and dental materials applications because of their superior mechanical properties and high corrosion resistance. However, the results in term of osseointegration has not always been satisfactory under altered metabolic bone conditions [1, 2]. The bone formation around an implant is a complex process and it is not fully understood. The surface morphology, roughness, chemical composition and coating of implant with selected materials, have been commonly investigated to increase the likelihood of osseointegration [3-8]. Among the methods used for the coating of biomaterials, the hydroxyapatite-coated prostheses have been generally fabricated by plasma spraying process [9].
In vivo evaluation constitutes a crucially important phase in the development of new implant materials. A wide variety of animal models have been used to evaluate the HA-coated samples, such as rats, rabbits and etc. For samples with potential load-bearing applications, a common practice is to place a number of small devices into animal at the intramedullary canals, or through cortical bone (transcortical) or subcutaneous site, and to evaluate the tissue adaptation to the material by both mechanical and histological techniques. Although it has been generally accepted that plasma-sprayed HA and CaP coatings improve bone strength and initial osseointegration rate, optimal coating properties, required to achieve maximal bone response are yet to be investigated [10, 11]. The aim of the present study was to investigate the materials characteristics and in vivo behavior of HA coatings prepared via plasma spraying process in our laboratories.
2. Materials and methods
2.1. Implant preparation and characterization
HA coatings with thickness varying from 25 to 75 μm were deposited by air plasma spraying on the Ti6Al4V bars with a diameter of 25 mm. Before thermal spraying, the Ti6Al4V samples were sand blasted in order to obtain a favorable surface for mechanical bonding in the coating/substrate interface. The roughness of substrate has an important role in the coating/substrate attachment [12]. The arithmetic average roughness parameter (Ra) of substrate surface used in this study was 2.3 μm. The phase characterization was performed by the X-ray diffraction method (Siemens D-500). After the surface preparation of the samples with 3 μm grade polishing, microstructural studies were conducted by scanning electron microscopy (Cambridge Stereoscan 360).
2.2. Experimental design for in vivo experiments
This study was performed in accordance with the University of Tehran Law on animal experiments. Fifteen adult male New Zealand rabbits 3.2±0.15 Kg body weight were submitted to dental bit drilling for unilateral implantation after exposure of the right medial femoral condyle. Acepromazine maleate (2% Karon Co. Iran) used 1 mg/kg.bw intramus-cularly as premedication and anesthesia was induced intramuscularly using 35 mg/Kg.bw ketamine hydrochloride(10% alfasan, Woerden-Holland) with combination of xylazine hydrochloride (2% alfasan Woerden-Holland) 5 mg/Kg.bw. After securing in the supine position of each rabbit, the right medial surface of stifle joint was shaved and prepared aseptically for operation. Skin incision given on the inner surface of medial condyle and after separation of connective tissues and fascia, a hole in size 3×4 mm was made in the all rabbit. A hole of 3 mm in diameter and 5 mm in depth transversely oriented was created unilaterally in the medial surface, and then the hole flushed and cooled with sterile 0.9% normal saline for removing bone debris.
The rabbits were subsequently divided into 3 groups of 5 animals per group. One piece of coated samples was inserted into each hole in group I. Group II contained the uncoated samples and no samples were used in the group III. Antibiotics (penicillin G procaine 40000 IU/kg IM every 12 h) dexamethasone 0.6 mg/Kg.b.w vitamin B. complex 0.2 mg/Kg.b.w and analgesic such as Tramadol hydrochloride (5 mg/Kg.b.w IM every 12 h) were administered for 3 post-operative days. The lateral and cranio-caudal radiographs were taken before and after implantation and on 15, 30, 45 and 60 days.
Related interpretations were done taking into consideration the extent of dislocation, new-bone formation, osteolysis and sclerosis. All rabbits were pharmacologically euthanized under general anesthesia by intravenous injection of thiopental sodium 10 %. At the experimental time of 60 days, the right femoral bones were explanted, excised and cleaned off soft tissue. The cubical segment of epiphysis part of each one was obtained using bone cutting system. All of the measurement carried out by blinded operators.
Figure 1. Cross sectional (left) and outer surface images (right) of plasma-sprayed HA coaing taken by SEM.
Figure 2. XRD pattern of plasma-sprayed HA coaing.
2.3. Histopathologic analysis
The bone segments, previously fixed in 10% neutral buffered formaldehyde solution were decalcified, embedded in paraffin and sectioned at 6 microns. The sections were obtained parallel (serial sections) and transverse to the long axis of the bar samples. These sections were stained with hematoxylin and Eosin (H and E) and Masson's trichrome methods for histopathology [13].
3. Results and discussion
The SEM images of lateral and outer surface of coating are shown in Figure 1. A double-featured, well-known porous and lamellar microstructure due to the plasma spraying parameters [14-16] was found in this work. However, the traces of discontinuity, microcracking and intact particles can be identified. The XRD analysis revealed that the principal crystalline phase of coating is hydroxypapatite (Figure 2). The presence of HA as the principal phase of plasma-sprayed HA coatings is reported elsewhere [14, 17]. The hardness of coatings and substrates was measured by Vickers indentation method. Absolute hardness values were 233 VHN for the substrate and 209 VHN for a 120 μm-thick coating. The difference between coating and substrate hardness values can be due to the material characteristics and the presence of porosity in the coating [18].
The initial attachment and subsequent proliferation of osteogenic cells, as well as their ability to form bone are the most important factors affecting the success of a bone implant [19]. Coating the implant with selected materials has been widely studied to improve tissue interaction [14, 15]. Concerning the in vivo behavior of our samples, no operative or postoperative complications were encountered. There were no visible signs of infection or other complications in these animals; all animals recovered from surgery and walked normally.
After sacrifice, the macroscopic evaluation revealed all coated (groups II) and uncoated (groups III) samples were well incorporated into the bone. The radiographic interpretation did not indicate any specific changes in the groups II and III (Figure 3). The histological examination of sections of all three groups revealed no evidence of infection, inflammation, foreign body reaction or necrosis of the tissue surrounding the samples. In control group, the holes were completely filled with dense connective tissue containing relatively parallel collagen bundles and immature bone trabecular fragments and unmineralized osteoid tissue. The bone trabecular structure was immature with many remodeling marks.
Figure 3. A HA-coated sample, indicating no local reaction and dislocation.
In uncoated group, around the samples there was not any macrophage or giant cell (Figure 4). The gap between bone and implant was filled with connective tissue containing immature bone trabeculae. Only at some places, a direct contact between bone and implant were occurred. In this group, the samples were covered by dense collagen fibers (Figure 4). Thin newly formed immature woven bone trabeculae were present parallelly to the implant surface. These trabeculae were lined with active osteoblasts. The new bone trabeculae were thinner and immature as compared to the group of coated samples. In coated group, the gap between the implant surface and drill wall was filled with mature connective tissue that extensively replaced with bone trabeculae and bridged the gap completely (Figures 5 and 6). However, a thin layer of HA remnants was present between the implant and connective tissue in some sites (Figures 5 and 6). The bone trabeculae structure was relatively mature with high density (mature lamellar bone). Bone formation was more significant where a well direct contact between coated implant and drill wall existed.
Figure 4. Light micrographs of tissue responses around sample at 60 days after implantation of uncoated titanium in femur of rabbit. Grafted material (a), dense connective tissue with irregular collagen fibers (b), immature woven bone trabeculae (c) were seen. The material was removed from the section. A: H&E×247; B: H&E × 395.
Figure 5. Light micrographs of tissue responses around material at 60 days after implantation of HA coated titanium in femur of rabbit. Grafted material (a), thin collagen fibers and HA remnant (b), trabeculae of lamellar bone (c), immature woven bone trabeculae (e). Bone marrow cavity (d) under the trabeculae was seen (H&E×110).
Figure 6. Light micrographs of tissue responses around material at 60 days after implantation of HA coated titanium in femur of rabbit. HA coated surface of implant (a), dense connective tissue with irregular collagen fibers (b), immature woven bone trabeculae (c) (H&E×282).
The bone-like apatite can be formed on the surface of titanium substrates in SBF environment [5-8] but for the in vivo conditions, the findings are controversial. It seems the uncoated titanium does not show any considerable bone-bonding properties for almost the same time after explantation used in this study [20, 21].
4. Conclusion
In this work, HA coatings were prepared by plasma spraying process and microstruc-tural properties of coatings and bone tissue response were investigated. The microstructure of coatings was porous and lamellar. Radiographic interpretation did not show osteolysis and inflammation or infection around samples in any groups. Histologic results revealed a higher level of osseointegra-tion for the HA-coated samples compared to the uncoated ones. Bone maturation was more significant at the bone-implant interface. Coating of titanium with hydroxyapatite led also to improve the maturation of new formed bone tissue. These results can be attributed to the presence of porous hydroxyapatite in the coated samples. Due to the high biocompat-ibility and osteoconduction of calcium phosphate materials, they have been widely used for different hard tissue applications such as HA-coated metallic implants and bone substitute materials [10, 11].
Acknowledgements
The authors wish to acknowledge the ENSAM (France), Materials and Energy Research Center (Iran) and Faculty of Veterinary Medicine and University of Tehran Research Council (Iran) for approval, financial and technical support of this research work.